Hostname: page-component-77c89778f8-7drxs Total loading time: 0 Render date: 2024-07-18T21:21:44.831Z Has data issue: false hasContentIssue false

Novel Features Incorporated in a New Automatic X-Ray Fluorescence Spectrometer

Published online by Cambridge University Press:  06 March 2019

Peter B. Whitcomb*
Affiliation:
Hilger & Watts, Inc., Chicago, Illinois
Get access

Abstract

A flat-crystal spectrometer is described that is fully evacuated to permit analysis of light elements. A uniselector in conjunction with a programing unit allows the crystal to be moved to any series of angles with stopping intervals as close as 0.01°. Detectors are a scintillation counter and a flow proportional counter driven at twice the angular velocity of the crystal to maintain them in correct position, A further detector is used to monitor the primary X-ray beam. Programing is achieved by means of a patchboard, and each board allows up to 20 elements to be measured at one time and in a sequence as desired, Pinal results appear in tabular forrn on an electric typewriter. Provisions are also provided for continuous scanning, the X-ray spectrum being plotted on a strip chart recorder. The sample is brought into position with a slide arranged so that the spectrometer may be kept continually evacuated and only a small pocket of air need be pumped. Alternately, helium may be used with resultant low consumption per sample.

Type
Research Article
Copyright
Copyright © International Centre for Diffraction Data 1960

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)