Hostname: page-component-77c89778f8-vsgnj Total loading time: 0 Render date: 2024-07-19T10:24:06.288Z Has data issue: false hasContentIssue false

The use of a newly developed pediatric arterial perfusion cannula for neonates

Published online by Cambridge University Press:  19 August 2008

Toshihide Asou*
Affiliation:
Department of Cardiovascular Surgery, Fukuoka Children's Hospital, Fukuoka, Japan
Hisataka Yasui
Affiliation:
Division of Cardiac Surgery, Research Institute of Angiocardiology, Kyushu University, Fukuoka, Japan
Hideaki Kado
Affiliation:
Department of Cardiovascular Surgery, Fukuoka Children's Hospital, Fukuoka, Japan
Yutaka Imoto
Affiliation:
Division of Cardiac Surgery, Research Institute of Angiocardiology, Kyushu University, Fukuoka, Japan
Yuichi Shiokawa
Affiliation:
Department of Cardiovascular Surgery, Fukuoka Children's Hospital, Fukuoka, Japan
Kouji Fukae
Affiliation:
Department of Cardiovascular Surgery, Fukuoka Children's Hospital, Fukuoka, Japan
Ryuji Tominaga
Affiliation:
Department of Cardiovascular Surgery, Fukuoka Children's Hospital, Fukuoka, Japan
Yoshito Kawachi
Affiliation:
Division of Cardiac Surgery, Research Institute of Angiocardiology, Kyushu University, Fukuoka, Japan
*
Dr Toshihde Asou, Department of Cardiac Surgery, Rumah Sakit Jantung ‘Harapan Kita’, National Cardiac Center, J l. S. Parman Kav. 87, Jakarta 11420, Indonesia Fax: (21) 568 4130

Abstract

We report our experience using an originally designed thin walled metal-tipped arterial cannula for neonatal open heart surgery. Its external diameter is 2.1 mm and its internal diameter is 1.7 mm, it has a high internal – external diameter ratio. It was used on 36 patients, who were either neonates or small infants. In 32 of the patients, it was inserted into the ascending aorta for systemic perfusion. In the other 4, it was inserted into the brachiocephalic artery for selective cerebral perfusion during repair of the aortic arch. The relationship between the pressure drop and the bypass flow was obtained for the 7 most recently treated patients. The pressure drop at a bypass flow of 500 ml/min was 150 mmHg. Due to the high internal – external diameter ratio of this cannula, and its excellent flow characteristics, we consider it to be particularly useful in neonatal open heart surgery, while it also allows for successful selective cerebral perfusion during repair of the aortic arch in neonates.

Type
Original Articles
Copyright
Copyright © Cambridge University Press 1997

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

Elliot, MJ. Cannulation for cardiopulmonary bypass for repair of congenital heart disease. In: Jonas, RA, Elliot, MJ (ed). Cardiopulmonary Bypass in Neonates, Infants and Young Children. But-terworth–Heinemann, Oxford, 1994, pp127139.Google Scholar
Yasui, H, Kado, H, Yonenaga, K, Kawasaki, S, Shiokawa, Y, Kohno, H, Tominaga, R, Kawachi, Y, Tokunaga, K. Revised technique of cardiopulmonary bypass in one-stage repair of interrupted aortic arch complex. Ann Thorac Surg 1993:55:11661171.Google ScholarPubMed
Van Meurs, KP, Mikesell, GT, Seale, WR, Short, BL, Rivera, O. Maximum blood flow rates for arterial cannulae used in neonatal ECMO. ASAIO Transactions 1990;36:M679681.Google ScholarPubMed