Most scanning electron microscopy is performed at low magnification; applications utilising the large depth of field nature of the SEM image rather than the high resolution aspect. Some environmental SEMs have a particular limitation in that the field of view is restricted by a pressure limiting aperture (PLA) at the beam entry point of the specimen chamber. With the original ElectroScan design, the E-3 model ESEM utilised a 500 urn aperture which gave a very limited field of view (∼550um diameter at a 10mm working distance [WD]). An increase of aperture size to ∼lmm provided an improved but still unsatisfactory field of view. The simplest option to increase the field of view in an ESEM was noted to be a movement of the pressure and field, limiting aperture back towards the scan coils1. This approach increased the field of view to ∼2mm, at a 10mm WD. A commercial low magnification device extended this concept and indicated the attainment of conventional fields of view.
- Type
- Environmental SEM
- Information
- Microscopy and Microanalysis , Volume 3 , Issue S2: Proceedings: Microscopy & Microanalysis '97, Microscopy Society of America 55th Annual Meeting, Microbeam Analysis Society 31st Annual Meeting, Histochemical Society 48th Annual Meeting, Cleveland, Ohio, August 10-14, 1997 , August 1997 , pp. 1193 - 1194