Using a Polycapillary Optic as a Spatial Filter to Improve Micro Xray Analysis in Low-Vacuum and Environmental SEM Systems

Published online by Cambridge University Press:  02 July 2020

Ning Gao
Affiliation:
aX-Ray Optical Systems, Inc.,30 Corporate Circle, Albany, NY, 12203
David Rohdeb
Affiliation:
ThermoNORAN, 2551 W. Beltline Highway, Middleton, WI, 53562-2697
Get access

Abstract

An inevitable consequence of the presence of the gas in the sample chamber of a low-vacuum scanning electron microscope (LV-SEM) and environmental SEM (ESEM) is the electron beam broadening due to the scattering in the gas. The electron broadening has a large impact on x-ray analysis because the fluorescent characteristic x rays generated far from the center of the electron probe form a high background, which reduces the detection sensitivity of x-ray analysis and degrades the x-ray image contrast. We report in this paper of using a polycapillary focusing x-ray optic between the sample and the energy-dispersive spectrometer as a spatial filter to filter out unwanted x-rays generated far from the specimen. As a result, the x-ray image contrast and the detection sensitivity of the system were notably improved.

A polycapillary focusing optic collects a large solid angle of x rays from an x-ray source of small area at its input focus, guide them through the curved channels by multiple external total reflections, and focus them to the output focus.

Type
Quantitative X-Ray Microanalysis in the Microprobe, in the SEM and in The ESEM:Theory and Practice (Organized by R. Gauvin and E. Lifshin)
Copyright
Copyright © Microscopy Society of America 2001

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

References:

1.Bolon, R. B., X-ray microanalysis in the ESEM, Microbeam Analysis, 199200 (1991)Google Scholar
2.Gao, N., Yu. Ponomarev, I., Xiao, Q.F., Gibson, W. M., Carpenter, D. A., Appl. Phys. Lett., 69, 1529(1996)CrossRefGoogle Scholar