Analysis of Agglomerate and Slag by Energy Dispersive X-Ray Fluorescence
Published online by Cambridge University Press: 06 March 2019
Extract
A rapid analytical procedure for simultaneous determination of Ca, Mn and Fe in agglomerate and slag has been developed.
A 100 mCi annular source of 238Pu has been used for excitation and Si(Li) detection system with ADC coupled with minicomputer for data acquisition and processing. A modified Heinrich-Rasberry procedure has been applied for analytical results evaluation. The measurement of each sample consumes about 100 s and the data processing time is 20 s approximately. The peaks were integrated in a fixed number of channels and linear background was subtracted. Seven standard samples both of slag and agglomerate have been used for system calibration.