Hostname: page-component-78c5997874-t5tsf Total loading time: 0 Render date: 2024-11-01T20:17:35.743Z Has data issue: false hasContentIssue false

Bone Mineral (Apatite) Assay by Photon Scattering - A Survey of Sources and Experimental Evaluation

Published online by Cambridge University Press:  06 March 2019

Luther E. Preuss
Affiliation:
Edsel B. Ford Research Institute, Physics Department Detroit, Michigan 48202
Dennis G. Piper
Affiliation:
Edsel B. Ford Research Institute, Physics Department Detroit, Michigan 48202
Claudius Bugenis
Affiliation:
Edsel B. Ford Research Institute, Physics Department Detroit, Michigan 48202
Get access

Abstract

Current methods of measuring bone mineral content In vitro are either inaccurate or measure density in non-intuitive units. A recently developed system overcomes these difficulties by utilizing the Compton scattering of photons from bone. Two sources of monoenergetic photons with related properties are required. The range includes energetic x-rays and low energy gamma rays. This study analyzes a number of the possible nuclide source combinations, and reports experimental results accomplished with a 153Gd-170-Tm combination. In vitro measurement of the density of ox bones by this method agreed with Archimedean measurements within three percent.

Type
Research Article
Copyright
Copyright © International Centre for Diffraction Data 1974

Access options

Get access to the full version of this content by using one of the access options below. (Log in options will check for institutional or personal access. Content may require purchase if you do not have access.)

References

1. Lachman, E., “Osteoporosis: The Potentialities and Limitations of Its Roentgenologic Diagnosis,” American Journal of Roentgenology 74, 712-15 (1955).Google Scholar
2. Garnett, E, S., Kennett, T. J., Kenyon, D. B., and Webber, C. E., “A Photon Scattering Technique for the Measurement of Absolute Bone Density in Man,” Radiology 106, 209-12 (1973).Google Scholar
3. Clarke, R. I. and Van Dyk, G., “A New Method for Measurement of Bone Mineral Content using both Transmitted and Scattered Beams of Gamma-rays,” Physics in Medicine and Biology 18, 532-9 (1973).Google Scholar
4. Johns, H. E. and Cunningham, J. R., The Physics of Radiology, p. 176, C. C. Thomas (1959).Google Scholar
5. Tipson, I. H., “Stable Element Metabolism,” in Health Physics Division Annual Progress Report, USAEC Report ORNL-4446, p. 300, Oak Ridge National Laboratory (1969).Google Scholar
6. McMaster, W. H., Del Grande, N. Kerr, Mallett, J. H., and Hubbell, J. H., Compilation of X-ray Cross Sections, UCRL- 50174, Sec. II Rev. 1, University of California, Livermore (1969).Google Scholar
7. Morgan, K. Z. and Turner, J. E., Principles of Radiation Protection, p. 21, John Wiley and Sons, Inc. (1967).Google Scholar