Published online by Cambridge University Press: 31 January 2005
Purpose: This paper describes our experiences of implementing systematic in vivo dosimetry at the Norfolk and Norwich Hospital and reviews the results of 2,254 entrance dose measurements made over a 17-month period.
Methods and materials: Entrance dose measurements using p-type diodes were performed on all new planned patients. The calibration procedure and correction factors applied are described. A 4% tolerance was applied.
Results: The results of all measurements indicated a small mean deviation from expected entrance dose of 10.77% and a standard deviation of 2.85%. 16.7% of all measurements exceeded the 4% tolerance with 9.2% exceeding a 5% level. The estimated overall errors for 578 treatments were calculated using the weighted averages of all beams. A narrower SD of 1.96% combined with only 4.8% of all treatments exceeding a 4% tolerance show that large deviations from a single field do not always translate into significant overall errors.
Conclusions: Global dosimetric accuracy was within clinically acceptable limits and variations between measured and expected doses were mainly attributable to factors affecting diode reading. A number of errors in calculating deviations and the inconsistent application of the protocol suggest the need for interfacing the diode system with software control.