Tools for examining subglacial bed deformation

Published online by Cambridge University Press:  20 January 2017

Erik Blake
Affiliation:
Department of Geophysics and Astronomy, University of British Columbia, Vancouver, British Columbia V6T 1Z4, Canada
Garry K. C. Clarke
Affiliation:
Department of Geophysics and Astronomy, University of British Columbia, Vancouver, British Columbia V6T 1Z4, Canada
Marc C. Gérin
Affiliation:
Department of Geophysics and Astronomy, University of British Columbia, Vancouver, British Columbia V6T 1Z4, Canada
Rights & Permissions [Opens in a new window]

Abstract

Deformation beneath soft-bedded glaciers may be a physical mechanism that contributes to flow instabilities such as surging. If the role of bed deformation is to be understood, a rheological description is required, but the development of a rheology is hampered by a lack of in situ stress and strain measurements. In this paper, we describe four techniques for measuring subglacial strain. Three of these give continuous strain measurements, a capability that permits calculation of instantaneous strain rates and allows comparison of strain data with other time series. To demonstrate the practicability of the techniques, sample results from three summers of experimentation beneath Trapridge Glacier. Yukon Territory, are presented. The data show that subglacial strain rate can vary in amplitude and polarity on an hourly time-scale,and that the instantaneous strain rate can exceed the mean strain rale by an order of magnitude. Observed negative strain rates suggest extrusive flow within basal sediments.

Information

Type
Research Article
Copyright
Copyright © International Glaciological Society 1992
Figure 0

Fig. 1. Schematic diagram (not to scale) of the borehole percussion hammer used to insert flexible strain instruments into the soft bed of Trapridge Glacier. The tool is sufficiently narrow that a figure drawn to scale would hide details.

Figure 1

Fig. 2. Tracings, from photographs, of Trapridge Glacier bed casts on a 10 cm by 10 cm grid, (a) A bed cast catalyzed and withdrawn immediately after insertion. (b) A bed cast left in the deforming layer for 92 h before removal (the tip was lost upon withdrawal). The dashed line indicates the upper boundary of the undisturbed bed into which the bedcasts were hammered; the dotted line indicates the estimated location of the ice-bed interface.

Figure 2

Fig. 3. Results from the 1987 rubber-rod experiment. The vertical scale is in relative strain units, as the device was uncalibrated. The solid line represents strain in the direction of ice flow. The dotted line represents cross-flow strain.

Figure 3

Fig. 4. Schematic diagram of the electrolytic tilt cell and electronics used in the 1988 experiment. The cell is 57 mm long and 16 mm in diameter. An a.c. excited bridge is formed between the cell and two reference resistors. As the cell is tilled, the lateral electrodes dip in and out of the electrolyte and their resistance with respect to the central electrode changes. The bridge imbalance is measured to give the degree and polarity of tilt. For dual axis-tilt sensitivity, a second pair of lateral electrodes is mounted at right-angles to the first pair and Ike central electrode is shared by both circuits.

Figure 4

Fig. 5. The 23 d record from the 1988 electrolytic tilt-sensor string. The upper, middle and lower cells are labelled (a), (b) and (c), respectively. Solid lines indicate tilt from vertical (left scale): dotted lines indicate the azimuth of the tilt with respect to the internal coordinates of the cell (right scale). The records from the two lower cells indicate that they were within a deformation Zone that sometimes experienced zero and negative strain rates. The arrows indicate these times. The cartoon below the graphs shows the position of the individual cells within the deforming layer at specific times. Ice movement is to the right.

Figure 5

Fig. 6. Schematic diagram of the leaf-spring tilt cell used in the 1989 experiment. The cell is 47 mm long and 16 mm in diameter. As the cell is tilted, the pendulum masses bend the two leaf springs mounted at right-angles to each other. The leaf springs are only sensitive to bending in one direction, so the two pendula respond to tilt along different axes. The bending is measured by strain gauges bonded to the surfaces of the leaf springs.

Figure 6

Fig. 7. Sample results from the 1989 leaf-spring tilt-sensor experiment. Tilt records for each of the three tilt cells have been differentiated to give strain rate parallel to ice flow. Strain rates normal to ice flow have been omitted for clarity, but are of comparable magnitude.